FDAAugust 23, 2019device
Pharos Excimer Laser, Model: EX-308 - Product Usage: The Pharos EX-308 Excimer Laser System is designed to treat the skin disorders psoriasis, vitiligo, atopic dermatitis, and leukoderma using phototherapy. The excimer laser system produces collimated 308 nm radiation, within the invisible region...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause overtreatment, which could result in strong erythema, and possibly cause blisters.
What to do
FDA enforcement status: Terminated
Brands named
ra medical systems
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAGSMS Incorporated, NIACIN EXTENDED-RELEASE TABLETS, USP, 1,000 MG, 90 tablets, Rx only, Manufactured by Kremers Urban Pharmaceuticals Inc., a subsidiary of Lannett, Inc., Seymour, IN 47274, Packaged by GSMS Incorporated, Camarillo, CA 93012. NDC 51407-268-90.2026-05-29
- FDAPhilips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # ...2026-05-18
- FDAPhilips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Elev...2026-05-18
- CPSCGenerac Power Systems Expands Recall of Generac and DR Power Electric Start Pressure Washers Due to Risk of Serious Injury or Death from Carbon Monoxide Hazard; Includes Additional Models2026-05-14
- FDANicotine Transdermal System Patch, 14 mg, 14 patches per box, Aveva, Distributed by: RUGBY LABORATORIES, Indianapolis, IN 46268, www.majorpharmaceuticals.com, NDC 0536-5895-88.2026-05-14
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAMerlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software2026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →