FDAFebruary 2, 2015device

Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Medical Axial Torque Limiting Driver is a reusable surgical instrument designed to limit the magnitude of applied (input) torque transferred to an external (output) device or fastener in the clockwise direction. The device ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.

What to do

FDA enforcement status: Terminated

Brands named

bradshaw medicalbradshaw