FDAMay 1, 2023device

Trilogy Evo O2 and LifeVent EVO2, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, LA2100X15B, IA2100X15B, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, EU2100X15B, EU2100X19, EE2100X15B, and SP2100X26B

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

What to do

FDA enforcement status: Ongoing

Brands named

philips respironicsphilips

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.