FDAApril 3, 2018device

Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRP Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood prod...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.

What to do

FDA enforcement status: Terminated

Brands named

baxter healthcarebaxter