FDAApril 11, 2023device

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

What to do

FDA enforcement status: Ongoing

Brands named

medtronic navigationmedtronic

UPCs

00763000253165007630003067620076300029992700763000299972007630005172670076300051727400763000517335