FDAApril 11, 2023device
StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
What to do
FDA enforcement status: Ongoing
Brands named
medtronic navigationmedtronic
UPCs
00763000253165007630003067620076300029992700763000299972007630005172670076300051727400763000517335
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAMedtronic O-arm O2 Imaging System. Model Number: BI70002000.2026-05-19
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
- FDAOctopus 4 Tissue Stabilizer, Model 294002026-05-12
- FDAMedtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter2026-04-21
- FDAMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94965; Cardiopulmonary bypass vascular catheter2026-04-21