FDAMarch 26, 2021device

PIC070 Arterial Blood Sampler PICO70 - Product Usage: Preheparinized, electrolyte-balanced, arterial blood sampler product line for collection of arterial samples for pH, blood gas, oximetry, electrolyte and metabolite analysis. Model Number/Product Description: 956-518 PICO70 w/o needle; 956-519...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Radiometer has become aware of an issue with PICO70 Arterial Blood Sampler. The issue relates to the sterile barrier system for the product. During internal stress testing on PICO70 Arterial Blood Samplers a breach in the sterile barrier system was observed in one sample. Risk for the patient: The described error is considered to have a remote possibility of resulting in bacterial bloodstream infection. The bloodstream infection may be asymptomatic but may also progress to sepsis or life-threatening septic shock. Immunosuppressed patients are at particular risk. Due to issues (breaches) with the sterile barrier system for the Arterial Blood Sampler.

What to do

FDA enforcement status: Terminated

Brands named

radiometer medical apsradiometerradiometer medical