FDAMarch 19, 2025device

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.

What to do

FDA enforcement status: Ongoing

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare