FDANovember 2, 2017device

NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)064381472...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.

What to do

FDA enforcement status: Terminated

Brands named

perkinelmer life and analytical sciences wallac oyperkinelmerperkinelmer life