FDAMarch 3, 2017device
Dimension Vista 500 Intelligent Lab System and Dimension Vista 1500 Intelligent Lab System, analyzer, chemistry (photometric, discrete), for clinical use Product Usage: The Siemens Healthcare Diagnostics Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytic...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Siemens Healthcare Diagnostics has confirmed that in isolated cases when ECREA is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of greater than 15 percent in the ECREA result.
What to do
FDA enforcement status: Terminated
Brands named
siemens healthcare diagnosticssiemenssiemens healthcare
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAChlorthalidone Tablets, USP, 25 mg, [100 or 1000] Tablets pr bottle, Rx only, Manufactured by: Inventia Healthcare Limited, Additional Ambernath, M.I.D.C., Ambernath (East) - 421506, INDIA. Distributed by: Risiong Pharma Holdings, Inc., East Brunswick, NJ 08816. NDC 100-tablet bottle: 64980-599-0...2026-06-05
- FDASiemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay2026-05-07
- CPSCBayer Recalls 6 mL Size Afrin Original Nasal Spray Bottles Due to Risk of Serious Injury or Illness from Child Poisoning; Violates Mandatory Standard for Child-Resistant Packaging2026-04-30
- FDAdirectCHECK ACT-LR whole blood quality control, level 2 Model/Catalog Number: DCGLR-22026-04-30
- FDABrand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.2026-04-22
- FDABioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.2026-04-22
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;2026-04-02
- FDABiofire Spotfire Respiratory/Sore Throat Panel REF: 4234852026-04-01
Own something like this?