FDAApril 20, 2015device

Dimension Vista Systems LOCI Thyroid Stimulating Hormone; the TSH method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone in human serum and plasma on the Dimension Vista System.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well.

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.