Positron Emission Tomograpy and Computed Tomography System, Model: uMI 550, with software version R001.3.0.0.750505, Rx Only, CE 0197, UDI: (01)06971576832026 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, di...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential sporadic software bugs in R001.3.0.0.750505 software version may cause an issue with the PET acquisition raw data causing reconstruction problems during PET/CT scanning and could possibly cause rescans of patients with additional dose.

What to do

FDA enforcement status: Terminated