Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The Alinity c Cuvettes are used on the Alinity c system processing module. The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest proces...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.
What to do
FDA enforcement status: Terminated
Brands named
abbott gmbh co kgabbottabbott gmbh