FDAFebruary 16, 2026device

Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THOR...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lpmedlinemedline industries

UPCs

101984596063974019845960639810195327479473401953274794741019845923727040198459237271108843898577004088438985770110889942212491408899422124921019845924569540198459245696