FDAMarch 20, 2024device
ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-017-12
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
What to do
FDA enforcement status: Ongoing
Brands named
zimmer
UPCs
0088902408853547235701712
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 1100402402026-05-20
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 1100402412026-05-20
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 1100404442026-05-20
- FDAOff-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 1100404762026-05-20
- FDAOff-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 1100402422026-05-20
- FDABrand Name: Persona Revision Product Name: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Model/Catalog Number: 42-5572-066-10 Software Version: N/A Product Description: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Size 9, 9+ 10mm Thicknes Component: N/A2026-05-11
- FDABrand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A2026-03-30
- FDABrand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A2025-12-24