FDAMarch 20, 2024device

ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-06

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

What to do

FDA enforcement status: Ongoing

Brands named

zimmer

UPCs

0088902405587200235701706

Recall history