FDAOctober 19, 2017device
BEQ-TOP 50903 Custom ECC Pack-Pedriatric , Custom Tubing Kit, Catalog No. 709000078
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
What to do
FDA enforcement status: Terminated
Brands named
datascope
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABattery Charging Station; Model: 0998-00-0802;2026-02-06
- FDACS100 IABP. Software Version CS100 IABP Q.01.2026-01-23
- FDACS300 IABP. Software Version CS300 IABP C.01.2026-01-23
- FDACardiosave Rescue. Intra-Aortic Balloon Pump system.2025-11-04
- FDACardiosave Hybrid. Intra-Aortic Balloon Pump system.2025-11-04
- FDACardiosave Hybrid2025-10-23
- FDACardiosave Rescue2025-10-23
- FDAMEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-012024-09-17
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.