Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.

What to do

FDA enforcement status: Ongoing

Brands named

philips medical systems clevelandphilipsphilips medical

UPCs

0088483809908100848381030540088483810296500884838102972008848381029890088483810319100884838103566