FDAFebruary 4, 2020device

HEMO-7 Hemospray Endoscopic Hemostat

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.

What to do

FDA enforcement status: Terminated

Brands named

wilson cook medicalwilsonwilson cook