FDAJune 23, 2022device

INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.

What to do

FDA enforcement status: Ongoing