MiniMed 620G Insulin Pump (MMT-1750)
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAMedtronic O-arm O2 Imaging System. Model Number: BI70002000.2026-05-19
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
- FDAOctopus 4 Tissue Stabilizer, Model 294002026-05-12
- FDAMedtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter2026-04-21
- FDAMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94965; Cardiopulmonary bypass vascular catheter2026-04-21
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →