FDAFebruary 24, 2017device

BrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studi...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.

What to do

FDA enforcement status: Terminated

Brands named

philips medical systems clevelandphilipsphilips medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
BrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studi... — Recall Details · AllClear