FDAApril 6, 2020device

System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremiti...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical