FDAMarch 15, 2021device

MiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL;¿ PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL;¿ KIT MMT-1896WW MM780G 6.5W BLE SF MG;¿ PUMP MMT-1884XC 780G V6.5 CLIN US MG;¿ PUMP MMT-1885XC 780G V6.5 CLIN MM;¿ PUMP MMT-1886XC 780G V6.5 CLIN MG;¿ ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.

What to do

FDA enforcement status: Ongoing

Brands named

medtronic minimedmedtronic

UPCs

007630003842580076300038426500763000384272007630003842890076300045561300763000421939007630004219460076300044143200763000441449007630004542960076300041154100763000411374

Recall history