FDAApril 5, 2018device

Thoratec HeartMate 3 LVAS Implant Kit, Rx Only, Catalog # 106524US, 106524, 10652INT - LVAS KIT, HM 3 Indicated for providing short-term hemodynamic support in patients with advanced refractory left ventricular heart failure.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death.

What to do

FDA enforcement status: Terminated

Brands named

abbott

UPCs

00813024013297