Welch Allyn ProBP 2400 Digital Blood Pressure Device #2400, REF 901096, Rx ONLY, Made in China --- Distributed by Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153 --- Microlife Corporation 9F, 431, RuiGuang Road, NeiHu, Taipei, 11492, Taiwan, R.O.C. --- Microlife AG Espenstra...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
What to do
FDA enforcement status: Terminated
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280...2021-09-03
- FDABraun ThermoScan PRO 6000 Ear Thermometer-indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment. Model: 60002021-07-20
- FDAGS777 Wall Transformer Model: 77710 Intended Use: The GS777 Wall Transformer is intended for use in ambulatory and acute care settings.2021-04-22
- FDAProBP 3400 Digital Blood Pressure Device ((handheld, desk mount, wall mount models) Model: 34XFHT-B Intended: ProBP 3400 automatically measures systolic and diastolic pressure (excluding neonates) and pulse rate, as well as calculates Mean Arterial Pressure (MAP).2021-04-22
- FDAAM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.2021-03-31
- FDASurveyor S12/S19 Bedside Patient Monitor, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.2021-03-31
- FDAELI 280 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.2021-03-31
- FDAELI 380 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.2021-03-31