FDAMarch 18, 2024device

GE Healthcare Allia IGS 7, Interventional fluoroscopic x-ray systems

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.

What to do

FDA enforcement status: Ongoing

Brands named

ge medical systems scs