FDAFebruary 27, 2020device

Atellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare

UPCs

00630414002163

Recall history