FDAApril 9, 2024device

VITROS Immunodiagnostic Products Free T3 Reagent Pack-IVD is for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 131 5589

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

T3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.

What to do

FDA enforcement status: Ongoing

Brands named

quidelortho

UPCs

10758750008797

Recall history