FDAMarch 22, 2017device

The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

What to do

FDA enforcement status: Terminated