FDAApril 9, 2024device

The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VT

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

What to do

FDA enforcement status: Ongoing

Brands named

sophysa