FDAMay 3, 2024device

Achieva 3.0T Model Number (REF): (1) 781345; (2) 781344; (3) 781278; (4) 781277; (5) 781177

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

What to do

FDA enforcement status: Ongoing

Brands named

philips north americaphilipsphilips north

UPCs

00884838004139