Vacutainer Tubes Vacutainer tubes are intended for clinical laboratory hematology studies including but not limited to lead level testing, FEP level testing, and CBC testing.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Avid Medical Inc has received a complaint was received from an end-user indicating the vacutainer component 519700 (BD: 367863) expiring 12/16 was discovered inside of a kit labelled with an expiration date after 12/16. Upon receipt of complaint, AVID Medical promptly initiated an internal investigation and root cause analysis. Investigation found mixed component lots were contained in product 519700 component case. This occurrence resulted in the incorrect lot number and expiration being attributed to all work orders produced using the affected component. This issue was immediately contained and corrected. A full inventory analysis was conducted reviewing all vacutainer component inventories. The issue was found to affect a total of five vacutainer components.

What to do

FDA enforcement status: Terminated