FDAMay 10, 2019device

Leadwire sets with snap ends: REF/Catalog Number Description 2106381-001 ECG Leadwire set, 5-lead, snap, AHA, 74 cm/ 29 in 2106381-002 ECG Leadwire set, 5-lead, snap, AHA, 130 cm/ 51 in 2106381-003 ECG Leadwire set, 5-lead, snap, IEC, 74 cm/ 29 in 2106381-004 ECG Leadwire set, 5-lead, snap, IEC, ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

What to do

FDA enforcement status: Terminated

Brands named

ge healthcare