FDAMay 10, 2019device

Replacement ECG leadwires with snap ends REF/Catalog Number Description 2106386-101 Replacement ECG Leadwire, snap, LL, AHA, 130 cm/ 51 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

What to do

FDA enforcement status: Terminated

Brands named

ge healthcare

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?