FDAMarch 15, 2016device
STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by person...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.
What to do
FDA enforcement status: Terminated
Brands named
ventlab
Recall history
No related federal recalls on record for this brand yet.