Dimension Vista 500 Intelligent Lab System running on software versions V.3.6.1 SP1 or V.3.6.2, Device listing # D011374, All serial numbers and lots are affected. The Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mix...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample aspirations when processing tubes in Sample Racks that are front loaded on the Dimension Vista System.

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Dimension Vista 500 Intelligent Lab System running on software versions V.3.6.1 SP1 or V.3.6.2, Device listing # D011374, All serial numbers and lots are affected. The Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mix... — Recall Details · AllClear