Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print Version The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABrand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER Model/Catalog Number: 7730IALS Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: no2025-12-15
- FDABrand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700REA Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No2025-12-15
- FDABrand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700RE Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No2025-12-15
- FDABrand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Catalog Number: 7730ALS Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No2025-12-15
- FDAAdult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F2025-05-01
- FDAVentlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Horizon Adult Resuscitator REF #s: 9122HZ, HZI9000, HZI9108, 9106HZ, 9110HZ, 9120HZ, 9111HZ, 9100HZ, 9105HZ, 9108HZ, 9006HZ, 9002HZ; 2. Horizon Infant Resuscitator REF #s: 6112HZ,...2023-10-11
- FDAVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2, AF5140MB-D6, AF5140MB-F, AF1140M...2023-10-11
- FDAVentlab, LLC RescuMed manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. RescuMed Adult Resuscitator REF #s: BVM510-F, BVMB510-10, BVMB510-F, BVMB510S-F; 2. RescuMed Infant Resuscitator REF #s: BVMB710 & BVMB710-F; 3. RescuMed Pediatric Resuscitator R...2023-10-11
Own something like this?