FDAMay 15, 2025device

Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print Version The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.

What to do

FDA enforcement status: Ongoing

Brands named

sunmed

UPCs

1088948358897030889483588974

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.