FDAMay 16, 2019device

Centricity PACS Foundation Centricity PACS software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing of digital images and data acquired from diagnostic imaging devices.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.

What to do

FDA enforcement status: Terminated