FDAMay 7, 2025device

Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.

What to do

FDA enforcement status: Ongoing

Brands named

natus neurology dba excel tech ltd xlteknatusnatus neurology

UPCs

00382830010825

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.