FDAMay 7, 2015device

LDL-Cholesterol plus gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

What to do

FDA enforcement status: Terminated

Brands named

roche diagnostics operationsrocheroche diagnostics

UPCs

054016821900303886632205171369190

Recall history