FDAMay 12, 2025device

Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

What to do

FDA enforcement status: Ongoing

Brands named

dexcom

UPCs

00386270002198003862700025180038627000252500386270004291003862700042600038627000227304595986293080003862700020200038627000212900386270003638

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.