FDAMay 2, 2016device

LCP Dia-Meta Volar Distal Radius Plate 15 H Shaft/LT-Sterile; Indicated for fractures, osteotomies, and non-unions of the radius and other small bones.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Product incorrectly packaged. The labels on the outside of the box and on the LCP Dia-Meta Volar Distal Radius Plate inside the box are correct; however, the preprinted box which contains the LCP-Dia-Meta Volar Distal Radius Plate is incorrect. The box is preprinted for a 4.5 mm LCP Condylar Plate - Distal Femur.

What to do

FDA enforcement status: Terminated

Brands named

synthes usa productssynthessynthes usa