FDAJune 12, 2019device

AXIOM Artis BA, Model Number 5904656

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical