INLIVEN CRT-P, Model Numbers: a) V284 b) V285 c) W274 d) W275 (added 7/15/2021)
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population
What to do
FDA enforcement status: Ongoing
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOrca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.2026-05-12
- FDAIntera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device2026-05-11
- FDABoston Scientific Model SC-4318, Clik X Anchor REF: M365SC431802026-04-29
- FDABoston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10-3M,CSK Electrode, STAINLESS STEEL, 3M Cable, 10 cm x 28 ga (0.36 mm); REF: CSK-TC15, CSK Electrode, STAINLESS STEEL, 15 cm x 28 ga (0.36 mm); REF: CSK-TC15-3M, CSK Electrode, STA...2026-04-29
- FDABoston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.0175 in); REF:TCN-20, TCN Electrode, NITINOL, 20 cm x 0.45 mm (0.0175 in) REF: T...2026-04-29
- FDALUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.52026-03-30
- FDALUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.52026-03-30
- FDABoston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.2026-03-19