P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84" MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA (c) PROMATT PLUS W/SIDEWALLS, 42X80, MATERIAL NUMBER PMS4280NZBA (d) PROMATT PLUS 36IN MATRESS, MATERIAL NUMBER PMP3680NZBA (e) PROMATT WITH SIDEWALLS...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.

What to do

FDA enforcement status: Terminated

Brands named

joerns healthcarejoerns