FDASeptember 23, 2013device
Volcano s5i/s5ix/s5iz/CORE Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5i -806300-020 ; s5ix/s5iz - 807400-001; CORE - 400-0 I 00.02. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.
What to do
FDA enforcement status: Terminated
Brands named
volcano
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAVolcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.2025-06-13
- FDAPhilips OmniWire Pressure guide wire REF 89185J PN 3000002528912022-05-16
- FDAVolcano Visions PV. 035 Digital IVUS Catheter REF 88901, PN: 300005384002 Rx Only, Sterile EO, CE 27972022-05-13
- FDAVolcano Visions PV. 035 Digital IVUS Catheter REF 81234, PN: 300007367341 , Rx Only, Sterile EO, CE 27972022-05-13
- FDAMM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems2022-02-03
- FDAIntraSight Mobile IVUS Systems, Model: 797415, Part numbers: 300002018101, 3000020181022022-01-24
- FDASyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x2022-01-19
- FDAPhilips Volcano FFR software used in the following systems connected to the McKesson, Schwarzer, and some Philips Xper Hemodynamic Systems (HMS): (1) Volcano Imaging System s5i, Part #807400001; (2) s5ix Imaging System with s5iz PC Workstation Refurbished, Part#807400001 R and 807401200; (3) s5i ...2018-11-07