FDAMay 29, 2019device

Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

What to do

FDA enforcement status: Terminated

Brands named

zyno medicalzyno

UPCs

00814371020204

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.