FDAMarch 22, 2017device

Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descending thoracic aorta (not including dissections) having vascular anatomy suitable for endovascular repair including: " I...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There have been five reports of graft thrombosis/occlusion during global commercial use, each following treatment for BTAI. One case resulted in patient death, and three cases resulted in reintervention.

What to do

FDA enforcement status: Terminated

Brands named

cook medicalcook

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descending thoracic aorta (not including dissections) having vascular anatomy suitable for endovascular repair including: " I... — Recall Details · AllClear