FDAMay 25, 2016device

LIFEPAK CR Plus and/or LIFEPAK EXPRESS are Automatic External Defibrillator(s) and are non-wearable. LIFEPAK CR Plus and LIFEPAK EXPRESS defibrillators are intended for use on patients in cardiac arrest.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

LIFEPAK CR Plus Automated External Defibrillators (AED) or LIFEPAK EXPRESS AED may fail to initiate voice prompts when the ON/OFF button is pressed and the lid is opened due to an internal component (reed switch) that can intermittently become fixed in the closed position. A defibrillator in this condition will fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated.

What to do

FDA enforcement status: Terminated

Brands named

physio controlphysiophysio control