FDAJune 10, 2025device
VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via tran...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
What to do
FDA enforcement status: Ongoing
Brands named
boston scientificboston
UPCs
0068544702030100685447019251
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOrca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.2026-05-12
- FDAIntera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device2026-05-11
- FDABoston Scientific Model SC-4318, Clik X Anchor REF: M365SC431802026-04-29
- FDABoston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10-3M,CSK Electrode, STAINLESS STEEL, 3M Cable, 10 cm x 28 ga (0.36 mm); REF: CSK-TC15, CSK Electrode, STAINLESS STEEL, 15 cm x 28 ga (0.36 mm); REF: CSK-TC15-3M, CSK Electrode, STA...2026-04-29
- FDABoston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.0175 in); REF:TCN-20, TCN Electrode, NITINOL, 20 cm x 0.45 mm (0.0175 in) REF: T...2026-04-29
- FDALUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.52026-03-30
- FDALUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.52026-03-30
- FDABoston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.2026-03-19